ملخص
During a 1-year period, transcatheter nonsurgical closures of patent ductus arteriosus (PDA) using the Rashkind PDA Occluder System were performed in 6 children (mean age 4 years; mean weight 14.75 kg). A 12-mm device was successfully implanted transvenously in 5 children with the narrowest ductal diameter o<4 mm. Similarly, a 17 mm device was successfully implanted transvenously in 1 child with a ductal diameter of < 4 mm. There were neither immediate nor late complications in any of the cases. Complete closure was achieved 15 min after the procedure in 4 patients with the 12 mm device. Color Doppler echo 24 h after implantation confirmed complete closure in all 4 patients but demonstrated persisting trivial shunt in the other 2 cases: 1 with a 12-mm device and the child with the 17 mm device. An evaluation 14 days after device insertion documented complete occlusion in these remaining cases. We were 100% successful in our attempts at device implantation without any complication or residual shunt. Our early experience confirms that, in selected children, PDA can be safely and successfully closed nonsurgically using the Rashkind double umbrella device.
| اللغة الأصلية | English |
|---|---|
| الصفحات (من إلى) | 137-145 |
| عدد الصفحات | 9 |
| دورية | Medical Principles and Practice |
| مستوى الصوت | 5 |
| رقم الإصدار | 3 |
| المعرِّفات الرقمية للأشياء | |
| حالة النشر | Published - 1996 |
بصمة
أدرس بدقة موضوعات البحث “Nonsurgical transcatheter closure of patent ductus arteriosus using rashkind double umbrella device: Initial experience in kuwait and review of the literature?'. فهما يشكلان معًا بصمة فريدة.قم بذكر هذا
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver