A system to investigate and adjust profile pattern of computed tomography dose index along the longitudinal-axis

  • Choirul Anam*
  • , Riska Amilia
  • , Ariij Naufal
  • , Yanurita Dwihapsari
  • , Geoff Dougherty
  • *Corresponding author for this work
  • Department of Physics
  • Universitas Diponegoro
  • Department of Physics
  • Institut Teknologi Sepuluh Nopember
  • California State University Channel Islands

Research output: Contribution to journal › Article › peer-review

Abstract

The purpose of this study to develop software to extract and investigate the profiles of the tube current and volume computed tomography dose index (CTDIvol) along the longitudinal axis (z-axis). The tube current and CTDIvol were extracted from the Digital Imaging and Communications in Medicine (DICOM) header of every image along the longitudinal axis. We evaluated the profiles of the tube current and CTDIvol from eight computed tomography (CT) scanners. If the CTDIvol did not fluctuate along the fluctuation of the tube currents, then the system will adjust the CTDIvol with tube currents. It is found that TCM is not always activated. If TCM is activated, the profiles of TCM vary from one scanner to another. The Siemens and Philip scanners have adjusted the CTDIvol profile with tube current, but the Toshiba scanner has not. By developed software, CTDIvol profile of the Toshiba can be easily adjusted. In conclusion, software to investigate the profile pattern of CTDIvol along the longitudinal axis has been successfully developed. The software is easy to use and works quickly. From this study, medical staff must be careful when using the CTDIvol along longitudinal axis contained in each DICOM header.

Original languageEnglish
Pages (from-to)1656-1662
Number of pages7
JournalInternational Journal of Public Health Science
Volume12
Issue number4
DOIs
StatePublished - Dec 2023
Externally publishedYes

Keywords

  • CT dose index
  • Indosect
  • Patient dose
  • Size-specific dose estimates
  • Water equivalent diameter (D)